J Interv Cardiol, 16:485-490, 2003
Paclitaxel-eluting stents: are they all equal? An analysis of six randomized controlled trials in de novo lesions of 3,319 patients.
Silber S
Cardiology Practice in the Dr. Müller Hospital, Munich,
Germany. silber@med.de
In Germany, four different drug eluting stents (DES) systems are
currently commercially available. Whereas sirolimus has been clinically
tested in only a single type of stent with a single type of coating
in only a single dose, paclitaxel has been tested on various stent
designs, in various dose densities, and in various release formulations
with or without a polymer carrier. Therefore, the question arises:
are all paclitaxel stents equally safe and effective? Six clinical
randomized trials investigated the safety and efficacy of paclitaxel-eluting
stents in patients with de-novo lesions: TAXUS-I (61 pats), TAXUS-II
(536 pats), ASPECT (177 pats), ELUTES (190 pats), DELIVER-I (1041
pats) and TAXUS-IV (1314 pats). In the TAXUS-series, paclitaxel
released from the stent was controlled by the Translute polymer.
In the other studies, however, no polymer carrier was used. In
TAXUS-I, II & IV, the dose density of 1 microg/mm2 significantly
reduced angiographic parameters of restenosis and improved clinical
outcomes. In ASPECT and ELUTES there was a dose-dependent effect
on angiographic parameters of restenosis with the best results
for a paclitaxel dose density of approximately 3.0 microg/mm2.
Clinical outcomes at 6 and 12 months, however, were not improved
in these studies without coating. The studies unanimously show
that the paclitaxel-eluting stents are safe, if clopidogrel is
added to ASA for 3 to 6 months. The safety of paclitaxel-eluting
stents is independent of the stent design, the dose density and
the presence or absence of a polymer carrier system. For paclitaxel-eluting
stents using a polymer carrier, the dose density of 1 microg/mm2
is highly effective, whereas for paclitaxel-eluting stents without
a polymer carrier, the minimal effective dose density is much
higher (3 microg/mm2). Despite their improvement of angiographic
parameters, paclitaxel-eluting stents without a polymer carrier
did not demonstrate a positive effect on clinical outcome. In
contrast, polymer-based paclitaxel elution produced significant
clinical benefit
Publication Types:
· Review
· Review, Academic
PMID: 14632945 [PubMed]